Improved Reconstruction Stability for Chemical Shift Encoded Hyperpolarized 13C Magnetic Resonance Spectroscopic Imaging Using K-T Spiral Acquisitions

MAGNETIC RESONANCE IN MEDICINE(2020)

Cited 1|Views9
No score
Abstract
PurposeA multiecho, field of view (FOV)‐oversampled k‐t spiral acquisition and direct iterative decomposition of water and fat with echo asymmetry and least‐squares estimation reconstruction is demonstrated to improve the stability of hyperpolarized 13C magnetic resonance spectroscopic imaging (MRSI) in the presence of signal ambiguities attributed to low‐SNR (signal‐to‐noise‐ratio) species, local uncertainties in metabolite peaks, and echo‐to‐echo signal inconsistencies.Theoryk‐t spiral acquisitions redistribute readout points to be more densely spaced radially in k‐space by acquiring an FOV and matrix that are oversampled by η. These more densely spaced spiral turns constitute effective intraspiral echoes and can supplement conventional interspiral echoes to improve spectral separation and reduce spectral cross‐talk to better resolve 13C‐labeled species for spectroscopic imaging.MethodsDigital simulations and imaging phantom experiments were performed for a range of interspiral echo spacings and η using multiecho, k‐t spiral acquisitions. Image spectral cross‐talk artifacts were evaluated both qualitatively and quantitatively as the percent error in measured metabolite ratios. In vivo murine experiments evaluated the feasibility of multiecho, k‐t spiral [1‐13C]pyruvate MRSI to reduce spectral cross‐talk for 3 scenarios of different expected reconstruction stability.ResultsDigital simulations and imaging phantom experiments both demonstrated reduced or comparable image spectral cross‐talk and percent errors in measured metabolite ratios with increasing η and better choices of echo spacings. In vivo images displayed markedly reduced spectral cross‐talk in lactate images acquired with η = 7 versus η = 1.ConclusionThe precision of hyperpolarized 13C metabolic imaging and quantification in the presence of low‐SNR species, local uncertainties in metabolite resonances, and echo‐to‐echo signal inconsistencies can be improved with the use of FOV‐oversampled k‐t spiral acquisitions.
More
Translated text
Key words
Perfusion Imaging,Magnetic Resonance Imaging,Metabolic Imaging,Hyperpolarization
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined