An extended core nanocoax pillar architecture for enhanced molecular detection.

L A D'Imperio, A E Valera,J R Naughton, M M Archibald,J M Merlo, T J Connolly,M J Burns,T C Chiles,M J Naughton

Biosensors and Bioelectronics(2019)

引用 1|浏览24
暂无评分
摘要
Biosensors that incorporate nanomaterials and nanofabrication techniques enable molecular detection of chemical and biological macromolecules with a high degree of specificity and ultrasensitivity. Here, we present a novel fabrication process that yields a nanostructure capable of detecting biological macromolecules. The extended core nanocoax (ECC) structure builds on a previously reported nanocoaxial-based sensor. The fabrication of the device incorporates an extended inner pillar, with controllable extension above the annulus and into the surrounding solution. This new design eliminates structural constraints inherent in the original nanocoax architecture. We also provide results demonstrating improvement in biosensing capability. Specifically, we show the capability of the new architecture to detect the B subunit of the Vibrio cholerae toxin at improved sensitivity (100 pg/ml) in comparison to optical enzyme-linked immunosorbant assay (1 ng/ml) and previously reported coaxial nanostructures (2 ng/ml).
更多
查看译文
关键词
Nanocoax,Nanofabrication,Extended-core,Biosensor,High-sensitivity,ELISA