Calibration of oscillometric non-invasive devices for monitoring blood pressure

METROLOGIA(2015)

Cited 5|Views5
No score
Abstract
Blood pressure is one of the most important vital signs used to monitor a patient's medical condition and is widely measured in hospitals and at home. Automatic, non-invasive blood pressure (NIBP) monitoring devices measure systolic and diastolic blood pressures from the analysis of cuff pressure oscillations caused by periodic variations of blood pressure in an artery. Currently, clinical validation by comparing them to the auscultatory reference has been used to verify the performance of NIBP devices. However, there are presently no calibration methods for NIBP devices. Here, we propose an SI-traceable calibration method for oscillometric NIBP devices. The calibration system generates pressure-pulses at pre-determined cuff pressures, and with pre-determined amplitude, to the device-under-test. The uncertainty of each pulse is analyzed and used for the calculation of blood pressure (BP) uncertainty. The maximum uncertainty for systolic and diastolic BP using the newly developed calibration system is (0.74 and 0.60) mmHg (k = 2) depending on the pressure and amplitude of each pulse, as well as the number of pulses applied. The present method can be used for calibration of oscillometric NIBP devices.
More
Translated text
Key words
blood pressure,calibration of non-invasive blood pressure (NIBP) monitoring devices,oscillometric method