Attenuation correction for the collimated gamma ray assay of cylindrical samples.

Applied Radiation and Isotopes(2015)

Cited 3|Views6
No score
Abstract
The Hybrid Monte Carlo (HMC) method developed earlier for attenuation correction of non-collimated samples [Agarwal et al., 2008, Nucl. Instrum. Methods A 597, 198], has been extended to the segmented gamma ray assay of cylindrical samples. The method has been validated both experimentally and theoretically. For experimental validation, the results of HMC calculation have been compared with the experimentally obtained attenuation correction factors. The HMC attenuation correction factors have also been compared with the results obtained from literature available near-field and far-field formulae at two sample-to-detector distances (10.3cm and 20.4cm). The method has been found to be valid at all sample-to-detector distances over a wide range of transmittance. On the other hand, the literature available near-field and far-field formulae have been found to work over a limited range of sample-to detector distances and transmittances. The HMC method has been further extended to circular collimated geometries where analytical formula for attenuation correction does not exist.
More
Translated text
Key words
Gamma-ray spectrometry,Self attenuation,Hybrid Monte Carlo method,Open ended Collimator,Far-field geometry,Near-field geometry