Extravascular Migration of the Catheter. A Rare Complication of the Totally Implantable Venous Device

Acta chirurgica Belgica(2002)

引用 23|浏览3
暂无评分
摘要
Since the first report of TIVD implantation, many authors have published immediate and post-operative complications. The Authors report the first case in which the catheter, implanted via the cephalic vein by cutdown technique, has migrated completely out of the vein. In order to avoid extravascular migration, during the surgical procedure, the catheter should be fixed at the vein with nonabsorbable sutures, taking care to avoid narrowing the catheter lumen. This case report discusses the possibility of catheter migration when a TIVD fails to function properly and considers how to avoid more disastrous effects.