Chrome Extension
WeChat Mini Program
Use on ChatGLM

Gas Dynamics In Co2 Angiography: In Vitro Evaluation In A Circulatory System Model

K Song,D Cho, K Shinn, E Charlton,K Cho

INVESTIGATIVE RADIOLOGY(1999)

Cited 18|Views4
No score
Abstract
RATIONALE AND OBJECTIVES. The use of carbon dioxide (CO2) as a vascular contrast agent has increased significantly since the introduction of digital subtraction angiography, To optimize the injection of CO2 for digital subtraction angiography, we evaluated the gas dispersion patterns from differing catheter designs, gas flow dynamics, and the influence of vessel size and inclination on luminal gas filling.METHODS. A circulatory system model was constructed and perfused with 36% glycerin solution at a rate of 1.08 to 1.13 liters per minutes (pulse rate 72 beats/minute, pressure 90-111 mm Hg). Fifty milliliters of CO2 was rapidly injected into the vascular tube of the model (diameter 6.4-15.9 mm) at an inclination of 0 degrees to 45 degrees via a catheter, and imaged digitally in a cross-table lateral projection. The dispersal patterns of gas bubbles from the halo, pigtail, and end-hole catheters were evaluated as well as the degree of luminal gas filling.RESULTS. The halo and end-hole catheters produced continuous gas flow with homogeneous density. The pigtail catheter produced smaller bubbles with inhomogeneous density. Luminal gas filling was incomplete, with a residual fluid level posteriorly regardless of the size and inclination of the tube. At 0 degrees inclination, gas filling was greater with the 6.4-mm tube than with the 15.9-mm tube. With an inclination of 0 degrees to 15 degrees, gas filling was significantly improved for the larger tubes. On dispersal, CO2 bubbles rapidly coalesced and moved forward along the anterior aspect of the tube. The frontal motion of the bubble was parabolic in configuration.CONCLUSIONS. The halo and end-hole catheters provide more homogeneous gas density than the pigtail catheter, Gas filling was incomplete regardless of catheter design, vessel size, or inclination. Inclination improves gas filling in vessels >12.7 mm in diameter.
More
Translated text
Key words
CO2, contrast media, CO2 angiography, catheters, pulsatile flow model
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined