Mixed venous oxygen saturation as a promising parameter for physiologic control of total artificial heart.

ASAIO journal (American Society for Artificial Internal Organs : 1992)(2000)

引用 9|浏览5
暂无评分
摘要
Mixed venous oxygen saturation (SvO2) has been proposed as one of the suitable parameters for physiologic control of a total artificial heart (TAH). To establish the practical application of SvO2, we investigated the response of cardiac output (CO) and SvO2 to step-loaded exercise. A normal calf was surgically equipped with an ultrasonic flowmeter probe and an oximetry catheter in the pulmonary artery to measure CO and SvO2, respectively. Three stage step treadmill exercise tests (1, 2, and 4 km/h) were performed three times. While CO increased from 8.9 L/min at preexercise level to 9.7, 10.2 and 11.4 L/min at 1, 2, and 4 km/h, respectively, SvO2 decreased from 59.6% to 56.8, 55.3, and 52.2%, respectively. There existed a linear correlation between the magnitude of changes in CO and SvO2. CO and SvO2 exhibited a similar course of change, expressing an inverted exponential curve. The time constant of SvO2 was from 19 to 35 seconds, whereas that of CO was from 21 to 39 seconds. We conclude that SvO2 changes in close association with CO during exercise and has good potential to be a parameter for physiologic control of a TAH, by reflecting the recipient's CO demand without conspicuous time delay.
更多
查看译文
关键词
Cardiac Imaging,Extracorporeal Membrane Oxygenation,Remote Ischemic Conditioning,Ventricular Assist Device,Cardiogenic Shock
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要