谷歌浏览器插件
订阅小程序
在清言上使用

TiO 2 sol-gel derived amperometric biosensor for H 2 O 2 on the electropolymerized phenazine methosulfate modified electrode

ANALYTICAL AND BIOANALYTICAL CHEMISTRY(2002)

引用 43|浏览15
暂无评分
摘要
. A novel hydrogen peroxide biosensor was developed based on the immobilization of horseradish peroxidase (HRP) in a TiO 2 sol-gel matrix on an electropolymerized phenazine methosulfate (PMS) modified electrode surface. Such membranes are of interest due to their high surface area, biological compatibility, and ease of fabrication. HRP entrapped in the TiO 2 matix was stable and retained its activity to a large extent. Cyclic voltammetry and amperometric measurements were employed to demonstrate the feasibility of electron transfer between immobilized HRP and the glassy carbon electrode via electropolymerized PMS. The influence of various experimental parameters such as operating potential, pH, temperature, and stability was investigated for optimum analytical performance. The biosensor provided a wide linear calibration range from 4.0×10 –6 M to 1.0×10 –3 M, with a detection limit of 8.0×10 –7 M at a signal-to-noise ratio of 3. The sensor retained 80% of its original activity after two months of operation.
更多
查看译文
关键词
TiO2 sol-gel Biosensor Phenazine methosulfate Horseradish peroxidase
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要