Feasibility of patient-specific quality assurance (PSQA) for real-time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces.

Journal of applied clinical medical physics(2024)

引用 0|浏览2
暂无评分
摘要
PURPOSE:To design a patient specific quality assurance (PSQA) process for the CyberKnife Synchrony system and quantify its dosimetric accuracy using a motion platform driven by patient tumor traces with rotation. METHODS:The CyberKnife Synchrony system was evaluated using a motion platform (MODUSQA) and a SRS MapCHECK phantom. The platform was programed to move in the superior-inferior (SI) direction based on tumor traces. The detector array housed by the StereoPhan was placed on the platform. Extra rotational angles in pitch (head down, 4.0° ± 0.15° or 1.2° ± 0.1°) were added to the moving phantom to examine robot capability of angle correction during delivery. A total of 15 Synchrony patients were performed SBRT PSQA on the moving phantom. All the results were benchmarked by the PSQA results based on static phantom. RESULTS:For smaller pitch angles, the mean gamma passing rates were 99.75% ± 0.87%, 98.63% ± 2.05%, and 93.11% ± 5.52%, for 3%/1 mm, 2%/1 mm, and 1%/1 mm, respectively. Large discrepancy in the passing rates was observed for different pitch angles due to limited angle correction by the robot. For larger pitch angles, the corresponding mean passing rates were dropped to 93.00% ± 10.91%, 88.05% ± 14.93%, and 80.38% ± 17.40%. When comparing with the static phantom, no significant statistic difference was observed for smaller pitch angles (p = 0.1 for 3%/1 mm), whereas a larger statistic difference was observed for larger pitch angles (p < 0.02 for all criteria). All the gamma passing rates were improved, if applying shift and rotation correction. CONCLUSIONS:The significance of this work is that it is the first study to benchmark PSQA for the CyberKnife Synchrony system using realistically moving phantoms with rotation. With reasonable delivery time, we found it may be feasible to perform PSQA for Synchrony patients with a realistic breathing pattern.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要