Outcomes of outflow graft stenting in HeartMate 3 left ventricular assist devices: A systematic review and individual patient data meta-analysis

ARTIFICIAL ORGANS(2024)

引用 0|浏览5
暂无评分
摘要
PurposeLVAD outflow graft stenosis continues to remain prevalent with a high complication rate. We sought to pool the existing evidence on indications, utilization patterns, and outcomes of transcatheter interventions for outflow graft stenosis in the HeartMate 3 LVAD. MethodsAn electronic search was performed to identify all studies in the English literature reporting on HeartMate 3 LVAD outflow graft stenting. Patient-level data were extracted for analysis. ResultsThirteen published reports and one unpublished case comprising a total of 28 patients were included. Median patient age was 68.5 years [Interquartile range: 58, 71] and 25.9% (7/27) were female. Dyspnea [60.7% (17/28)] was the most common presenting symptom. Low flow alarms were present in 60% (15/25) of patients. Findings included external compression [35.7% (10/28)], graft twist [21.4% (6/28)], graft twist and external compression [14.3% (4.28)], intraluminal thrombus [10.7% (3/28)], graft twist and intraluminal thrombus [3.6% (1/28)], and pseudoaneurysm of outflow graft [3.6% (1/28)]. Median time from LVAD implantation to stenting was 2.1 years [1.4, 3]. Immediate flow normalization after stenting was observed in 85.7% (24/28). The 30-day mortality was 12% (3/25). Overall mortality was 12% (3/25) at a median follow-up of 3.9 months [1, 17]. ConclusionOutflow graft stenting in the HeartMate 3 LVAD appears to be a reasonable treatment option for outflow graft stenosis, with low overall rates of complications and mortality. Further refinement of indications and approaches may improve outcomes.
更多
查看译文
关键词
balloon angioplasty,cannula obstruction,Heartmate 3,Heartmate III,HM3,LVAD,outflow graft obstruction,outflow graft stenosis,stenting,twisting
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要