Mid-infrared auto-correction on-chip waveguide gas sensor based on 2f/1f wavelength modulation spectroscopy

OPTICS LETTERS(2024)

引用 0|浏览3
暂无评分
摘要
Compared to the most commonly used on-chip direct absorption spectroscopy (DAS) gas detection technique, the second harmonic (2f) based on-chip wavelength modulation spectroscopy (WMS) proposed by our group has the faculty to suppress noise and improve performance, but the accuracy of 2f WMS is easily affected by optical power variation. A mid-infrared auto-correction on-chip gas sensor based on 2f /1f WMS was proposed for decreasing the influence of the variation of optical power. The limit of detection of methane (CH4) obtained by a chalcogenide waveguide with a length of 10 mm is 0.031%. Compared with the 2f WMS, the maximum relative concentration error of the auto-correction on-chip gas sensor was decreased by similar to 5.6 times. The measurement error is <= 2% in a temperature variation range of 30 degrees C. This auto-correction sensor without a complicated manual calibration is helpful to the high accuracy measurement for on-chip integrated gas sensing. (c) 2024 Optica Publishing Group