Improved Microelectrode Array Electrode Design for Heavy Metal Detection

Chemosensors(2024)

引用 0|浏览1
暂无评分
摘要
Traditional working electrodes are not sufficient to realize the low detection limit and wide detection range necessary for the detection of heavy metals. In this study, a microelectrode array electrode was proposed using a design scheme based on microelectromechanical systems that was optimized with finite element software. The working electrode adopted an innovative composite structure to realize the integrated design of the working and counter electrodes, which improved the system integration. Performance tests showed that the electrode realized the quantitative analysis of Cd(II), Pb(II), and Cu(II) with a low detection limit (0.1 μg/L) and a wide detection range (0.1–3000 μg/L). The electrode successfully measured the lead and copper ion concentrations in the Sanya River, including both seawater and freshwater environments. The experimental results demonstrate that the electrode exhibits excellent adaptability to environmental conditions and can be potentially applied for technical support in environmental monitoring and sewage treatment.
更多
查看译文
关键词
microelectrode array electrode,heavy metal,detection,dissolved voltammetry
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要