Test of 116CdWO4 and Li2MoO4 scintillating bolometers in the CROSS underground facility with upgraded detector suspension

A. Ahmine, I. C. Bandac,A. S. Barabash, V. Berest, L. Berge, J. M. Calvo-Mozota,P. Carniti, M. Chapellier,I. Dafinei, F. A. Danevich,T. Dixon, L. Dumoulin, F. Ferri,A. Giuliani, C. Gotti, P. Gras, D. L. Helis,A. Ianni, L. Imbert, H. Khalife, V. V. Kobychev, S. I. Konovalov, P. Loaiza, P. de Marcillac,S. Marnieros, C. A. Marrache-Kikuchi, M. Martinez, C. Nones,E. Olivieri, A. Ortiz de Solorzano, Y. Peinaud,G. Pessina, D. V. Poda,Th. Redon, Ph. Rosier,J. A. Scarpaci, V. I. Tretyak, V. I. Umatov, M. Velazquez, M. M. Zarytskyy, A. Zolotarova

JOURNAL OF INSTRUMENTATION(2023)

引用 0|浏览2
暂无评分
摘要
In preparation for the CROSS experiment to search for double-beta (2/3) decay of 100Mo, we upgraded a low-background facility at the Canfranc underground laboratory (Spain) by installing a magnetic damping-based detector suspension inside a pulse-tube cryostat. We tested the suspension in low-temperature (15 mK) measurements with two scintillating bolometers made of large-volume crystals of 116CdWO4 (reference of the set-up) and Li2MoO4 (R&D of the CLYMENE project in view of the CUPID 2/3 experiment), each coupled to a thin Ge bolometric light detector. Despite the evidence of a residual pulse-tube-induced noise, picked up by cabling, we achieved high performance with all tested devices. In particular, the energy resolution for 2615 keV y-ray measured with both 116CdWO4 and Li2MoO4 bolometers is -6 keV FWHM, among the best-reported results for thermal detectors based on such compounds representing a great interest to searches for 2/3 decay in 116Cd (Q-value is 2813 keV) and 100Mo (3034 keV) respectively. Highly efficient particle identification (a vs. /3/y) is achieved also with the scintillating bolometer made of a low light yield crystal (Li2MoO4), thanks to the low noise (10 eV RMS) exhibited by the light detector exploiting the Neganov-Trofimov-Luke effect for signal amplification. We also found a rather high level of radiopurity in the Li2MoO4 crystal; only traces of 210Po and 226Ra were detected (- 0.1 mBq/kg each), while the 228Th activity is expected to be at least an order of magnitude lower, as well as a 40K activity is below 6 mBq/kg.
更多
查看译文
关键词
Cryogenic detectors,Double-beta decay detectors,Solid state detectors,Particle identification methods
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要