Non-enzymatic paper-based analytical device for direct potentiometric detection of urine creatinine

Microchimica Acta(2024)

引用 0|浏览3
暂无评分
摘要
A paper-based analytical device (PAD) with an integrated composite electrode has been designed and fabricated for non-enzymatic creatinine sensing. Reduced graphene oxide (rGO) was employed to modify the PAD so that it could function as a solid-contact transducer. A new macrocyclic pyrido-hexapeptide derivative was made and used as a special ionophore in the creatinine membrane sensor. The synthesized PAD showed a detection limit of 1.0 µM (S/N = 3) and a potentiometric response towards creatinine throughout a log-linear range of 2.0 µM–10 mM ( R 2 = 0.9998). The sensor shows significant selectivity for a few related substances, including ephedrine, codeine, ketamine, caffeine, urea, urate, carbinoxamine, and dextromethorphan. It has been established that the testing method is appropriate for the direct potentiometric detection of creatinine in a variety of human urine sample types. When an indicating electrode and a reference electrode are put on the same flexible disposable, this lets applications with a small sample volume be done. For point-of-care creatinine measurement, the developed paper-based analytical equipment is a good choice because it is affordable, easily accessible, and self-pumping (especially when combined with potentiometric detection). Graphical Abstract
更多
查看译文
关键词
Paper-based analytical device,Potentiometric sensors,Carbon nanotube ink,Point-of-care,Urine creatinine