P094/306 The PEGASUS stent system for stent assisted coiling of cerebral aneurysms – a multicenter case series
Abstracts(2023)
摘要
Introduction
In this retrospective case series we evaluated stent-assisted coiling with a new low profile lasercut stent (pegasus, Phenox gmbh, Bochum, Germany)Aim of Study
To evaluate the performance of the pegasus stent system for aneurysm treatment.Methods
Patients treated with stent assisted coiling for intracranial aneurysm(s), with or without acute subarachnoid hemorrhage were included into the study. Clinical, imaging, procedural parameters and clinical and imaging follow up parameters were recorded and statistically evaluated.Results
54 Aneurysms in 53 patients from 6 centers were included into the study. Mean age was 57 years (+-11,68). 29 patients were treated electively, 24 acutely (23 SAH, one partially thrombosed aneurysm with ischemic events). Intracranial procedure related events were recorded in 1 (6,9%) elective patient and in 3 SAH (12,5%) patients. Postprocedural ischemic complications occurred in 3 (12,5%) of the SAH patients and in 1 of the elective patients (3,45%), the latter not changing the mRS of the patient. Overall aneurysm occlusion was Raymond Roy (RR) I in 36 cases, RR II in 9 and RR IIIa in 9 cases and all stents were patent on final DSA run. On the first control which was available for 23 patients after 147,7 days (+-59,6) RR I was achieved in 22, RR II in 1 patient.Conclusion
Stent assisted coiling with the pegasus stent system seems safe and effective.Disclosure of Interest
DL: Money paid to Institution by Phenox GmbH JK: Procotor for Phenox GmbH disclosures of the other authors will be delivered as soon as possible (too close to deadline)查看译文