Gain stabilization and adjustment method of multiple SiPM based gamma-ray detectors on GECAM

arXiv (Cornell University)(2021)

引用 0|浏览1
暂无评分
摘要
The Gravitational wave high-energy Electromagnetic Counterpart All-sky Monitor (GECAM) consists 25 SiPM based gamma-ray detectors(GRDs) on each payload. Although SiPM based GRD has merits of compact size and low bias-voltage, the drifts of the SiPM gain with temperature is a severe problem for GRD performance. An adaptive voltage supply source was designed to automatically adjust the SiPM bias voltage to keep the gain constant. This method has been proved to be effective both in ground experiment and in-flight test. The in-flight variation of the SiPM gain is within 2% by the temperature compensation. To reduce the non-uniformity of the GRD gain, an iterative bias voltage adjustment method is proposed and the uniformity is reduce from 17% to 0.3%. In this paper, the gain stabilization and adjustment method of multiple SiPM based GRDs are discussed in detail.