Integrated fluorescence excitation, collection, and filtering on a GaN waveguide chip

CHINESE OPTICS LETTERS(2023)

引用 0|浏览3
暂无评分
摘要
Fluorescence detection is widely used in biology and medicine, while the realization of on-chip fluorescence detection is vital for the portable and point-of-care test (POCT) application. In this Letter, we propose an efficient fluorescence excitation and collection system using an integrated GaN chip consisting of a slot waveguide and a one-dimensional photonic crystal (1D PC) waveguide. The slot waveguide is used to confine the excitation light for intense light-sample interaction, and the one-trip collection efficiency at the end of slot waveguide is up to 14.65%. More interestingly, due to the introduction of the 1D PC waveguide, the fluorescence signal is directly filtered out, and the excitation light is reflected to the slot waveguide for multiple excitations. Its transmittances for the designed exciting wavelength of 520 nm and the fluorescent wavelength of 612 nm are 0.2% and 85.4%, respectively. Finally, based on numerical analysis, the total fluorescence collection efficiency in our system amounts to 15.93%. It is the first time, to our knowledge, that the concept of an all-in-one-chip fluorescence detection system has been proposed, which paves the way for on-chip fluorescence excitation and collection, and may find potential applications of miniaturized and portable devices for biomedical fluorescence detection.
更多
查看译文
关键词
gan waveguide chip,integrated fluorescence excitation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要