Intraoperative use of electrical impedance spectroscopy for adenoid cystic carcinoma of the lacrimal gland: A case report

Clinical case reports(2023)

引用 0|浏览2
暂无评分
摘要
Key Clinical Message Electrical Impedimetric Tumor Detection System is a novel and promising tool for fast intraoperative tumor delineation and accurate safe margin detection in orbital tumors. Abstract Adenoid cystic carcinoma (ACC) is a rare malignant tumor of epithelial origin, typically arising from the salivary and lacrimal glands. ACC is notorious for recurrence and a high rate of morbidity and mortality despite therapy. We presented a 48‐year‐old male patient with lacrimal gland ACC of the right orbit who underwent radical tumor resection and adjuvant radiotherapy. We applied a new diagnostic method, the Electrical Impedimetric Tumor Detection System, during surgery and tested its performance to enhance the precision of tumor resection. Two months after surgery, he underwent external radiation of 58 Gy in 29 fractions. He showed no tumor recurrence or metastasis in the 1‐year follow‐up visits. ITDS showed a precision of tumor and margin detection consistent with histopathology results. This novel ITDS may be a reliable system for fast intraoperative tumor delineation and accurate, safe margin detection in orbital tumors.