An FFT-Based DC Offset Compensation and I/Q Imbalance Correction Algorithm for Bioradar Sensors

IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES(2024)

引用 0|浏览7
暂无评分
摘要
The challenge of noncontact presentation of human cardiopulmonary activity using a bioradar sensor is to linearly demodulate the Doppler cardiopulmonary diagram (DCD) signal from baseband signals. Arctangent demodulation can perform linear phase demodulation to obtain the DCD signal. However, the high-order harmonics and intermodulation terms (ITs) caused by the time-varying direct current (dc) offset and in-phase and quadrature-phase (I/Q) imbalance in the baseband signals significantly degrade the signal-to-noise ratio (SNR) of the Doppler heartbeat diagram (DHD) signal. In this work, a fast Fourier transform (FFT)-based algorithm is proposed to simultaneously perform time-varying dc offset compensation and I/Q imbalance correction without the need for an auxiliary device to improve the accuracy of the arctangent demodulation. The obtained results show that the SNRs of the algorithm-processed DHD signals are increased from 30.08 $\pm$ 2.41 to 68.88 $\pm$ 10.57 dB. In addition, the root mean square errors (RMSEs) of the C-C intervals of the DHD signals for eight subjects with respect to the J-J intervals of the ballistocardiogram (BCG) signals are 17.79 $\pm$ 2.72 ms (2.80% $\pm$ 0.43%), suggesting a promising potential of the DHD signal for noncontact biomedical applications.
更多
查看译文
关键词
Arctangent demodulation,bioradar sensor,direct current (dc) offset,Doppler heartbeat diagram (DHD),in-phase and quadrature-phase (I/Q) imbalance