In this work we describe a straightforward approach for creating a nanocomposite comprising multiwalled carbon nanotubes (MWCNTs) and titanium dioxide (TiO2) using the hydrothermal technique, which is then characterized by scanning electron microscope (SEM), energy-dispersive X-ray spectrometer (EDS), X-ray diffraction analysis (XRD), Fourier transform infrared spectroscopy (FTIR), and thermal gravimetric analysis (TGA) to assess its properties. Nafion is employed as a reticular agent for the nanocomposite on the glassy carbon electrode (GCE), creating the MWCNT/TiO2/Nafion/GCE system. The electrochemical behavior of the system was evaluated using cyclic voltammetry, revealing its remarkable electrocatalytic activity for detecting hydrogen peroxide in water. The developed sensor showcased a broad linear response range of 14.00 to 120.00 μM, with a low detection limit of 4.00 μM. This electrochemical sensor provides a simple and highly sensitive method for detecting hydrogen peroxide in aqueous solutions and shows promising potential for various real-world applications, particularly in H2O2 monitoring.