Simulation study of an x-ray sub-picosecond resolution detection system based on time-domain amplification.

Applied optics(2023)

引用 0|浏览12
暂无评分
摘要
This study proposes what we believe to be a novel x-ray detection system that achieves a temporal resolution of 930 fs with photorefractive and four-wave mixing effects. The system comprises two parts: a signal-conversion system and signal-acquisition system. The signal-conversion system is based on the photorefractive effect, which converts x-ray evolution into the variation of infrared interference intensity. The signal-conversion sensor consists of ultra-fast response LT-GaAs and a high-resolution interference cavity, achieving a resolution of  767 fs. The signal-acquisition system consists of a time-domain amplification system based on four-wave mixing and a high-resolution signal-recording system with a resolution of 21 ps, providing a temporal resolution of 525 fs.
更多
查看译文
关键词
resolution,x-ray,sub-picosecond,time-domain