Crystal Analyzer Based Multispectral Microtomography Using CCD-Sensor.

Sensors (Basel, Switzerland)(2023)

引用 0|浏览1
暂无评分
摘要
To solve the problems of spectral tomography, an X-ray optical scheme was proposed, using a crystal analyzer in Laue geometry between the sample and the detector, which allowed for the selection of predetermined pairs of wavelengths from the incident polychromatic radiation to obtain projection images. On a laboratory X-ray microtomography setup, an experiment was carried out for the first time where a mixture of micro-granules of sodium chloride NaCl, silver behenate AgC22H43O2, and lithium niobate LiNbO3 was used as a test sample to identify their spatial arrangement. The elements were chosen based on the presence of absorption edges in two of the elements in the energy range of the polychromatic spectrum of the probing radiation. The method of projection distortion correction was used to preprocess the obtained projections. To interpret the obtained reconstruction results, the segmentation method based on the analysis of joint histograms was used. This allowed us to identify each of the three substances. To compare the results obtained, additional "reference" tomographic measurements were performed: one in polychromatic and two in monochromatic (MoKα-, MoKβ-lines) modes. It took three times less time for the tomographic experiment with the crystal analyzer, while the reconstruction accuracy was comparable to that of the "reference" tomography.
更多
查看译文
关键词
X-ray imaging,tomography,K-edge subtraction imaging,reference tomography,crystal analyzer,spectral tomography,distortion correction algorithm