Lack of Consensus Between Measurements of Plasma Phylloquinone by Enzyme-Linked Immunosorbent assays and a Well-Validated High-Performance Liquid Chromatographic Method.

Current developments in nutrition(2023)

引用 0|浏览8
暂无评分
摘要
Enzyme-linked i2mmunosorbent assays (ELISAs) that measure circulating phylloquinone have become commercially available, but their validity is uncertain. The objective of this study was to compare plasma phylloquinone concentrations measured using two commercially available ELISAs with concentrations measured using a validated high-performance liquid chromatography (HPLC) assay in 108 samples obtained from participants in a depletion (∼10 mcg phylloquinone/d)-supplementation (∼500 mcg phylloquinone/d) study. The geometric mean of plasma phylloquinone measured with ELISA A was 0.70 nmol/L, 37% lower than that measured with HPLC. The mean of the ELISA B measures was 12.4 nmol/L, >700% higher than the HPLC measures. Plasma phylloquinone measured using HPLC was significantly lower during phylloquinone depletion than supplementation (0.4 ± 0.1 compared with 1.2 ± 0.2 nmol/L; < 0.001). Neither of the two ELISAs detected any significant difference in plasma phylloquinone concentrations between depletion and supplementation (ELISA A, = 0.76; ELISA B, = 0.29). These findings reinforce the need to validate plasma phylloquinone assays as they become available. 2023;x:xx.
更多
查看译文
关键词
phylloquinone,immunoassay,ELISA,quality assurance,vitamin K
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要