Testing the Use of Advanced Upper Limb Prostheses: Towards Quantifying the Movement Quality with Inertial-Magnetic Measurement Units

PROSTHESIS(2023)

引用 0|浏览12
暂无评分
摘要
Background: A thorough assessment of upper limb prostheses could help facilitate their transfer from scientific developments into the daily lives of users. Ideally, routine clinical testing would include assessments of upper limb function using motion-capturing technology. This is particularly relevant for the state-of-the-art upper limb prostheses. Methods: We designed a test based on an activity of daily life ("tray-task") which could be completed outside the laboratory, and developed a set of outcome measures aimed at characterizing the movement quality. For this purpose, kinematics of the thorax and the humerus were captured with an inertial-magnetic measurement unit (IMMU) motion-capture system. Six prosthesis users and ten able-bodied participants were recruited to test the feasibility of the proposed assessment procedure and to evaluate the outcome variables. Results: All participants completed the test either at home or in our lab. The prosthesis users needed more time to complete the task and showed a larger range of motion in the thoracic flexion and a smaller range of motion in the humeral elevation, compared to the able-bodied participants. Furthermore, the prosthesis users' movements were less smooth and characterized by less stable coordination patterns between the humerus and thorax. Conclusion: A new test method and associated outcome variables have been proposed.
更多
查看译文
关键词
arm prostheses,assessment,myoelectric control,motor coordination,outcomes