Spontaneous fracture of a titanium mesh cranioplasty implant in a child: A case report

WORLD JOURNAL OF CLINICAL CASES(2023)

引用 0|浏览2
暂无评分
摘要
BACKGROUNDTitanium mesh cranioplasty is often performed after decompressive craniectomy. Spontaneous fracture of the titanium prosthesis is an extremely rare postoperative complication. Here, we report a 10-year-old boy who presented with a spontaneous fracture of titanium mesh without antecedent head trauma.CASE SUMMARYA 10-year-old boy presented with a 1-wk history of a tender bulge over the left temporo-parieto-occipital scalp. He had undergone a temporo-parieto-occipital titanium mesh cranioplasty 26 mo previously. He denied antecedent head trauma. Computerized tomography disclosed a perpendicular fissure in the titanium mesh, suggesting a diagnosis of spontaneous titanium mesh fracture. He underwent a second temporo-parieto-occipital cranioplasty and made an uneventful recovery. Three-dimensional modeling and finite element analyses were used to explore potential risk factors of titanium mesh fracture.CONCLUSIONWe report a case of spontaneous fracture of a titanium mesh cranioplasty implant. The current case and literature review indicate that titanium mesh implants should be well-anchored to the base of bony defects to prevent fatigue-induced fractures.
更多
查看译文
关键词
Cranioplasty,Prosthesis fracture,Spontaneous fracture,Titanium,Case report