Evaluation of Image Quality and Detectability of Deep Learning Image Reconstruction (DLIR) Algorithm in Single- and Dual-energy CT

Journal of digital imaging(2023)

引用 0|浏览13
暂无评分
摘要
This study is aimed to evaluate effects of deep learning image reconstruction (DLIR) on image quality in single-energy CT (SECT) and dual-energy CT (DECT), in reference to adaptive statistical iterative reconstruction-V (ASIR-V). The Gammex 464 phantom was scanned in SECT and DECT modes at three dose levels (5, 10, and 20 mGy). Raw data were reconstructed using six algorithms: filtered back-projection (FBP), ASIR-V at 40% (AV-40) and 100% (AV-100) strength, and DLIR at low (DLIR-L), medium (DLIR-M), and high strength (DLIR-H), to generate SECT 120kVp images and DECT 120kVp-like images. Objective image quality metrics were computed, including noise power spectrum (NPS), task transfer function (TTF), and detectability index (d′). Subjective image quality evaluation, including image noise, texture, sharpness, overall quality, and low- and high-contrast detectability, was performed by six readers. DLIR-H reduced overall noise magnitudes from FBP by 55.2% in a more balanced way of low and high frequency ranges comparing to AV-40, and improved the TTF values at 50% for acrylic inserts by average percentages of 18.32%. Comparing to SECT 20 mGy AV-40 images, the DECT 10 mGy DLIR-H images showed 20.90% and 7.75% improvement in d′ for the small-object high-contrast and large-object low-contrast tasks, respectively. Subjective evaluation showed higher image quality and better detectability. At 50% of the radiation dose level, DECT with DLIR-H yields a gain in objective detectability index compared to full-dose AV-40 SECT images used in daily practice.
更多
查看译文
关键词
Deep learning,Image enhancement,Image reconstruction,Multidetector computed tomography,Radiation dosage
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要