Design and performance study of a proportional counter for low-energy X-rays

Meng Yu Li,Rui Zhao,Tao Xiong,Jin Jie Wu, Lin Fu, Tao Yu,Wei Feng Zhu, Xian Qiang Tang, Ping Zhou Lu, Bing Bing Qu

Radiation Physics and Chemistry(2023)

引用 0|浏览8
暂无评分
摘要
A cylindrical proportional counter based on a noble gas was designed to accurately measure low-energy X-ray radiation. The detection capability of the proportional counter with different gases and geometries was calculated through the Monte Carlo method. The optimal filling gas and applicable pipe diameter were determined according to the results of the Monte Carlo simulation. Then, energy calibration was performed at the monoenergetic X-ray radiation facility (MXRF) based on Bragg diffraction, and the energy calibration results were verified using the radionuclides 241Am and 129I. The relative deviation of the theoretical and measured values of the energy does not exceed 1.4%, and the energy resolution of the designed proportional counter is approximately 10%. It was proved that the linearity of the channel and energy of the proportional counter can be used for energy spectrum analysis, and the proportional counter has a good radionuclide recognition capability. Finally, the detection efficiency of the designed proportional counter was simulated and verified at multiple energy points. A detection efficiency calculation model was established through MCNP5, and the detection efficiency of the proportional counter was measured at the MXRF. The results of the experiment and simulation showed that the proportional counter could reach a highest detection efficiency of 92% in the low-energy region. The average deviation between the experiment and simulation was 1.63% at 10–34 keV and 3.90% at 34–75 keV. The proportional counter can be used for measurement and application of X-rays in the low-energy region.
更多
查看译文
关键词
Low-energy X-rays,Proportional counter,Energy calibration,Detection efficiency
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要