The comparison of diffusion tensor imaging in human hearts between 1.5 T and 3.0 T

BMC medical imaging(2023)

引用 0|浏览4
暂无评分
摘要
Background The aim was to compare the diffusion tensor imaging (DTI) indices derived from human hearts between 1.5 T and 3.0 T scanners. Additionally, the reproducibility of DTI indices was assessed between 1.5 T and 3.0 T scanners. Methods A total of 18 ex-vivo hearts were derived from patients who underwent heart transplantation. The DTI schemes were performed at 1.5 T and 3.0 T, respectively. Then, the same slices from each ex-vivo heart were selected for image analysis. The student’s t-test or Wilcoxon-rank test was used to compare the statistical differences. The agreement of DTI indices was mainly reported as the interclass correlation coefficient (ICC). Results No significant differences (all P > 0.05) were found in the DTI indices between 1.5 T and 3.0 T scanners. Interestingly, the ICC of all DTI indices was relatively lower with a low b-value. The reproducibility of the helix angle (HA) was relatively lower when compared to the other DTI indices. Conclusion The DTI indices of ex-vivo human hearts between 1.5 T and 3.0 T scanners had no significant differences. The consistency of DTI indices needed caution using a low b-value with different field strengths, and the relatively low reproducibility of HA should be considered.
更多
查看译文
关键词
Diffusion tensor imaging,Ex-vivo human heart,Cardiovascular magnetic resonance,Comparison