Association Between Device Type and Type IIIb Endoleaks Following Thoracic Endovascular Aortic Repair.

European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery(2023)

引用 4|浏览28
暂无评分
摘要
The Medtronic Valiant device met criteria for association with type IIIb endoleaks in the FDA's MAUDE database and met those criteria as early as 2013. A possible relationship between woven graft fabric and T3bEs is supported by the observation that all types of T3bE that occurred among a large number of TEVARs at the institution followed placement of grafts with sutured on woven fabric.