In this contribution we discuss the calibration and performance of the TOF-Wall detector of the Fragmentation-Of-Target (FOOT) experiment. A thin carbon target was irradiated with carbon ions of several energies ranging from 150 to 400 MeV/u. The charge spectra of the produced fragments were used to calibrate the energy response. The energy resolution for a single bar was about 6%. The time resolution between the two layers ranged from 42 (150 MeV/u) to 62 ps (400 MeV/u).