Attomole-Level Multiplexed Detection of Neurochemicals in Picoliter Droplets by On-Chip Nanoelectrospray Ionization Coupled to Mass Spectrometry

ANALYTICAL CHEMISTRY(2022)

引用 1|浏览15
暂无评分
摘要
While droplet microfluidics is becoming an effective tool for biomedical research, sensitive detection of droplet content is still challenging, especially for multiplexed analytes compartmentalized within ultrasmall droplets down to picoliter volumes. To enable such measurements, we demonstrate a silicon-based integrated microfluidic platform for multiplexed analysis of neurochemicals in picoliter droplets via nanoelectrospray ionization (nESI)-mass spectrometry (MS). An integrated silicon microfluidic chip comprising downscaled 7 mu m-radius channels, a compact T-junction for droplet generation, and an integrated nESI emitter tip is used for segmentation of analytes into picoliter compartments and their efficient delivery for subsequent MS detection. The developed system demonstrates effective detection of multiple neurochemicals encapsulated within oil-isolated plugs down to low picoliter volumes. Quantitative measurements for each neurochemical demonstrate limits of detection at the attomole level. Such results are promising for applications involving label-free and small-volume detection for monitoring a range of brain chemicals.
更多
查看译文
关键词
mass spectrometry,picoliter droplets,neurochemicals,attomole-level,on-chip