Electrochemical biosensor with electrokinetics-assisted molecular trapping for enhancing C-reactive protein detection

Biosensors and Bioelectronics(2022)

引用 16|浏览27
暂无评分
摘要
C-Reactive protein (CRP) is an essential biomarker relevant to various disease prognoses. Current biosensors require a significant amount of time for detecting CRP. To address this issue, this work proposes electrokinetic flow-assisted molecule trapping integrated with an impedance biosensor, where a driving signal in terms of a gated sine wave is provided to circularly arranged electrodes which detect proteins. To verify the biosensor's efficacy, protein aggregation on the electrode surface was evaluated through a fluorescence analysis and measurement of the electrochemical impedance spectrum (EIS). The fluorescence analysis with avidin showed that target samples largely accumulated on the electrode surface upon provision of the driving signal. The EIS measurement of CRP accumulation on the electrode surface further confirmed a significant electrokinetic phenomenon at the electrode/electrolyte interface. Even at the low CRP concentration of 10 pg/ml, the proposed device's sensitivity and reliability were as high as 3.92 pg/ml with a signal-to noise ratio (SNR) of ≥3, respectively. In addition, the protein detection time (without considering the preparation time) was minimized to as low as 90 s with the proposed device. This device's advantage is its minimal time consumption, and simple drop-analysis process flow; hence, it was used for monitoring clinical serum samples.
更多
查看译文
关键词
Electrochemical impedance spectroscopy (EIS),Dielectrophoresis,Electro-thermal convection,Preconcentration,C-Reactive protein
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要