How to Optimize Pedicle Screw Parameters for the Thoracic Spine? A Biomechanical and Finite Element Method Study

GLOBAL SPINE JOURNAL(2024)

引用 4|浏览1
暂无评分
摘要
Study Design: Pedicle screw study. Objective: The selection of pedicle screw parameters usually involves the surgeon's analysis of preoperative CT imaging along with anatomical landmarks and tactile examination. However, there is minimal consensus on a standardized guideline for selection methods on pedicle screws. We aimed to determine the effects of thoracic screw diameter to pedicle width on pullout strength determined by cortical bone purchase. Methods: Biomechanical study performed with human cadaveric thoracic vertebrae and experimentally validated three-dimensional finite element model instrumented with pedicle screws of various diameters. We used a variable (SD/PW) ratio to express the screw selection. We hypothesized a positive correlation between the pullout load determined by the bone purchase and the SD/PW. This relationship was first investigated in a validated finite element model considering bone purchase related to the strength of an upper thoracic vertebra. Then, the correlation to the entire spine is evaluated. Results: The failure load ranged from 371.3 to 1601.0 N, respectively, for 3 and 6 mm screws. The determinant coefficient was increased to R2=.421 when a linear relationship between pullout load and the SD/PW ratio was used. The peak loads of 1216 and 1288N were found for an SD/PW ratio of .83. Conclusion: We have found that the screw pullout load is more correlated to SD/PW than other pedicle measures for a maximized SD/PW ratio of .83. This particular value should be considered the upper limit of the indicated SD/PW ratio and a means to determine the optimal screw diameter to enhance pullout strength.
更多
查看译文
关键词
pedicle screws,FEA,breaches,thoracic spine,biomechanics,pullout strength
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要