Removing Ring Artefacts From Synchrotron Radiation-Based Hard X-Ray Tomography Data

DEVELOPMENTS IN X-RAY TOMOGRAPHY XI(2017)

引用 2|浏览7
暂无评分
摘要
In hard X-ray microtomography, ring artefacts regularly originate from improperly functioning pixel elements on the detector or from particles and scratches on the scintillator. We show that due to the high sensitivity of contemporary beamline setups further causes inducing inhomogeneities in the impinging wavefronts have to be considered. We propose in this study a method to correct the thereby induced failure of simple flatfield approaches. The main steps of the pipeline are (i) registration of the reference images with the radiographs (projections), (ii) integration of the flat-field corrected projection over the acquisition angle, (iii) high-pass filtering of the integrated projection, (iv) subtraction of filtered data from the flat-field corrected projections. The performance of the protocol is tested on data sets acquired at the beamline ID19 at ESRF using single distance phase tomography.
更多
查看译文
关键词
ring artefacts, tomography, 2D-2D registration, in-line tomography, high-pass filter, phase reconstruction