Surface Characterization of P-Type Point Contact Germanium Detectors

PARTICLES(2021)

引用 5|浏览9
暂无评分
摘要
P-type point contact (PPC) germanium detectors are used in rare event and low-background searches, including neutrinoless double beta (0 nu beta beta) decay, low-energy nuclear recoils, and coherent elastic neutrino-nucleus scattering. The detectors feature an excellent energy resolution, low detection thresholds down to the sub-keV range, and enhanced background rejection capabilities. However, due to their large passivated surface, separating the signal readout contact from the bias voltage electrode, PPC detectors are susceptible to surface effects such as charge build-up. A profound understanding of their response to surface events is essential. In this work, the response of a PPC detector to alpha and beta particles hitting the passivated surface was investigated in a multi-purpose scanning test stand. It is shown that the passivated surface can accumulate charges resulting in a radial-dependent degradation of the observed event energy. In addition, it is demonstrated that the pulse shapes of surface alpha events show characteristic features which can be used to discriminate against these events.
更多
查看译文
关键词
P-type point contact germanium detectors, alpha and beta surface backgrounds, neutrinoless double beta decay
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要