A Compton camera prototype for prompt gamma medical imaging

P. G. Thirolf,S. Aldawood,M. Boehmer,J. Bortfeldt, I. Castelhano,G. Dedes,F. Fiedler, R. Gernhaeuser,C. Golnik,S. Helmbrecht,F. Hueso-Gonzalez, H. V. D. Kolff,T. Kormoll, C. Lang, S. Liprandi, R. Lutter, T. Marinsek, L. Maier,G. Pausch,J. Petzoldt,K. Roemer,D. Schaart, K. Parodi

EPJ Web of Conferences(2016)

引用 20|浏览1
暂无评分
摘要
A Compton camera prototype for a position-sensitive detection of prompt gamma rays from proton-induced nuclear reactions is being developed in Garching. The detector system allows to track the Compton-scattered electrons. The camera consists of a monolithic LaBr3:Ce scintillation absorber crystal, read out by a multi-anode PMT, preceded by a stacked array of 6 double-sided silicon strip detectors acting as scatterers. The LaBr3:Ce crystal has been characterized with radioactive sources. Online commissioning measurements were performed with a pulsed deuteron beam at the Garching Tandem accelerator and with a clinical proton beam at the OncoRay facility in Dresden. The determination of the interaction point of the photons in the monolithic crystal was investigated.