Use of 2.1 MHz MRI scanner for brain imaging and its preliminary results in stroke.

Journal of magnetic resonance (San Diego, Calif. : 1997)(2020)

引用 32|浏览26
暂无评分
摘要
Cerebral stroke greatly contributes to death and disability rates in China and the whole world. Effective non-invasive imaging device for bedside monitoring of stroke is critically needed in clinically. This study developed a lightweight (350 kg) and low-footprint magnetic resonance imaging (MRI) system for brain imaging. Static magnetic field was built using an H-typed permanent magnet, which has 50.9 mT magnetic field strength (corresponding to 2.167 MHz proton Larmor frequency). Biplanar gradient coils were designed using the target field method based on dipole equivalent. Radio-frequency coils were optimized by particle swarm optimization. The 2 MHz MRI system was deployed in the Department of Neurology of hospital to test its performance in stroke imaging detection. Gradient recall echo and fast spin echo sequences were utilized to acquire T1- and T2-weighted MR images, respectively. Brain images of a healthy volunteer, a patient with hemorrhagic stroke, a patient of ischemic stroke, and a patient with ischemic stroke and images from 17-day long-term monitoring of hemorrhagic stroke were obtained with a 1.5 mm * 2.0 mm spatial resolution in plane, and a 10 mm thickness.