Objective: The use of a close-fitting roughly head-shaped volume co"/>

Optimization of a Close-Fitting Volume RF Coil for Brain Imaging at 6.5 mT Using Linear Programming

IEEE Transactions on Biomedical Engineering(2021)

引用 5|浏览4
暂无评分
摘要
Objective: The use of a close-fitting roughly head-shaped volume coil for MRI (magnetic resonance imaging) has the merit of improving the filling factor and thus the SNR (signal-to-noise ratio) from the brain. However, the surface of the RF coil follows that of the head which makes it difficult to determine an optimal coil winding pattern. We describe here a new method to optimize a head-shaped RF coil with the objective of maximizing its SNR and RF-magnetic-field homogeneity for operation at ultra-low magnetic field (6.5 mT, 276 kHz). Methods: The approach consists of FEM (finite-element-method) simulation and linear programing based optimization. Results: We have implemented the optimization and further studied the relationship between the design requirements and the performance of the RF coil. Finally, we constructed an optimal RF coil and scanned both a head-shaped phantom and a human subject. Conclusion: The method we outline here provide new insight into the conductor layout needed for magnetic optimization of structurally complex coils, especially when tradeoffs between competing attributes (SNR and homogeneity in this case) must be made.
更多
查看译文
关键词
Brain,Equipment Design,Humans,Magnetic Resonance Imaging,Neuroimaging,Phantoms, Imaging,Programming, Linear,Radio Waves