Model Reference Adaptive Control for Aortic Pressure Regulation in Ex Vivo Heart Perfusion

IEEE Transactions on Control Systems Technology(2021)

引用 4|浏览43
暂无评分
摘要
This brief proposes the first model reference adaptive control (MRAC) method for aortic pressure (AoP) regulation and maintaining the heart's physiological aerobic metabolism in ex vivo heart perfusion (EVHP). A mathematical model of the EVHP was established to describe the hemodynamic behavior of EVHP and quantify the changes in cardiac parameters. A reference model consisting of a virtual proportional-integral-derivative (PID) controller and the EVHP model was employed to generate the reference trajectory of AoP. An adaptation algorithm tunes the control parameters based on the reference model and the isolated heart. Experiments were conducted using large animal hearts (50 ± 5 kg porcine) to validate the adaptive controller's performance for stepwise and fast switching AoP references. The results confirmed the effectiveness of the proposed controller for regulating the AoP in isolated porcine hearts, in an accurate (mean error less than 2 mmHg) and fast (4-8 s of settling time) manner.
更多
查看译文
关键词
Automated organ perfusion,closed-loop control,Ex vivo heart,model reference adaptive control (MRAC)
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要