IMPROVEMENT OF GAMMA-RAY SUBTRACTION PROCEDURE FOR A CURRENT-MODE NEUTRON DETECTOR WITH A PAIR OF 6Li- AND 7Li-GLASS SCINTILLATORS.

RADIATION PROTECTION DOSIMETRY(2020)

引用 0|浏览16
暂无评分
摘要
A current-mode neutron detector with a pair of Li-6- and Li-7-glass scintillators has been developed to measure high-flux neutrons in a boron neutron capture therapy field. Neutrons are basically measured by subtracting gamma-ray component using current outputs from the Li-7-glass scintillator. In the present study, the difference in the gamma-ray sensitivity between the Li-6- and Li-7-glass scintillators and the neutron sensitivity for the Li-7-glass scintillator due to the Li-6 contamination were also considered to improve the gamma-ray subtraction precision. The gamma-ray subtraction procedure was experimentally investigated in thermal neutron fields with Cf-252 and Am-241-Be neutron sources, which have different gamma-ray intensities per unit neutron fluence. A linear relation between neutron fluence and current output was obtained for the neutron detector in the two types of thermal neutron fields with different gamma-ray intensities. It was found that the gamma-ray subtraction procedure is useful for current-mode neutron detectors.