Approaches to Cough Peak Flow Measurement With Duchenne Muscular Dystrophy.

RESPIRATORY CARE(2018)

引用 10|浏览9
暂无评分
摘要
BACKGROUND: Manually-assisted coughing and mechanical insufflation-exsufflation (MI-E) are commonly used in patients with Duchenne muscular dystrophy (DMD). Few studies have compared cough peak flow (CPF) with manually-assisted coughing to other methods, such as MI-E + manually-assisted coughing. In addition, few studies have reported the reliability of the measured CPF values. This study aimed to compare CPF with different cough-assistance methods and to examine the reliability of CPF data. METHODS: The study included 12 subjects with DMD (mean age, 34 +/- 8 y). CPF, CPF + manually-assisted coughing (assisted CPF), maximum insufflation capacity (MIC) + CPF (MIC-CPF), MIC + manually-assisted coughing (MIC+assisted CPF), MI-E (MI-E-CPF), and MI-E + assisted CPF were measured. A spirometer was used to compare CPF measurements obtained with each of the cough-assist techniques. The reliability of the measured CPF values was analyzed using Bland-Altman analysis. RESULTS: CPF was 59 +/- 34 L/min, assisted CPF was 113 +/- 32 L/min, MIC-CPF was 170 +/- 30 L/min, MIC+assisted CPF was 224 +/- 62 L/min, MI-E-CPF was 199 +/- 40 L/min, and MI-E + assisted CPF was 240 +/- 38 L/min. A fixed and proportional bias was found in the CPF measurements made with the peak flow meter and the spirometer. The average 95% CI in the difference between peak flow meter, MI-E, and CPF obtained using the spirometer were -7.45 to -1.95 and -1.45 to 4.95, respectively. Test for correlation was r = 0.54 (P < .001) for CPF (peak flow meter) and CPF (spirometer) and r = 0.17 (P = .17) in CPF (MI-E) and CPF (spirometer), respectively. CONCLUSION: MI-E + assisted CPF was the highest. The CPF measured with the peak flow meter suggested underestimation.
更多
查看译文
关键词
manually-assisted coughing,mechanical insufflation-exsufflation,cough peak flow,Duchenne muscular dystrophy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要