National Registry Data and Record Linkage to Inform Postmarket Surveillance of Prosthetic Aortic Valve Models Over 15 Years

JAMA INTERNAL MEDICINE(2017)

引用 28|浏览21
暂无评分
摘要
IMPORTANCE Postmarket evidence generation for medical devices is important yet limited for prosthetic aortic valve devices in the United Kingdom. OBJECTIVE To identify prosthetic aortic valve models that display unexpected patterns of mortality or reintervention using routinely collected national registry data and record linkage. DESIGN, SETTING, AND PARTICIPANTS This observational study used data from all National Health Service and private hospitals in England and Wales that submit data to the National Adult Cardiac Surgery Audit (NACSA). All patients undergoing first-time elective and urgent aortic valve replacement surgery (with or without coronary artery bypass grafting) with a biological (n = 15 series) or mechanical (n = 10 series) prosthetic valve from 5 primary suppliers, and satisfying prespecified data quality criteria (n = 43 782 biological; n = 11 084 mechanical) between 1998 and 2013 were included. Valves were classified into series of related models. Outcome tracking was performed using multifaceted record linkage. The median follow-up was 4.1 years (maximum, 15.3 years). Cox proportional hazards regression with random effects (frailty models) were used to model valve effects on the outcomes, with and without adjustment for preoperative and intraoperative covariates. MAIN OUTCOMES AND MEASURES Time to all-cause mortality or aortic valve reintervention (surgical or transcatheter). There were 13 104 deaths and 723 reinterventions during follow-up. RESULTS Of 79 345 isolated aortic valve replacement procedures with or without coronary artery bypass grafting, 54 866 were analyzed. Biological valve implantation rates increased from 59% in 1998 and 1999 to 86% in 2012 and 2013. Two series of valves associated with significantly increased hazard of death or reintervention were identified (first series: frailty, 1.18; 95% prediction interval [PI], 1.06-1.32 and second series: frailty, 1.19; 95% PI, 1.09-1.31). These results were robust to covariate adjustment and sensitivity analyses. There were 3 prosthetic valves with a significant reduction in hazard (valve 1: frailty, 0.88; 95% PI, 0.80-0.96; valve 2: frailty, 0.88; 95% PI, 0.80-0.96; and valve 3: frailty, 0.88; 95% PI, 0.78-0.98). CONCLUSIONS AND RELEVANCE Meaningful evidence from the analysis of routinely collected registry data can inform postmarket surveillance of medical devices. Although the findings are associated with a number of caveats, 2 specific biological aortic valve series identified in this study may warrant further investigation.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要