Parallel radiofrequency transmission at 3 tesla to improve safety in bilateral implanted wires in a heterogeneous model.

MAGNETIC RESONANCE IN MEDICINE(2017)

引用 50|浏览19
暂无评分
摘要
PurposeElongated implanted conductors can interact with the radiofrequency (RF) transmission field during MRI, posing safety concerns of excessive heating in patients with deep brain stimulators. A technique using parallel RF transmission (pTx) is evaluated on an anthropomorphic heterogeneous model with bilateral and unilateral curved wires. MethodsAmplitude and phase were optimized by simulation to minimize heating surrounding the implanted wires and to minimize B1+ inhomogeneity for four-channel and eight-channel pTx in a heterogeneous model. MRI experiments were conducted in an equivalent test phantom created from a common digital mesh file. ResultsIn four-channel pTx, maximum local specific absorption rate (SAR) was reduced in both unilateral and bilateral wires by 47% and 59%, respectively, but with compromised B1+ homogeneity. Optimized eight-channel pTx substantially reduced local SAR compared with birdcage coil RF excitation in both unilateral and bilateral wires (reduction of maximum local SAR of 79% and 87%, respectively). B1+ inhomogeneity was limited to 17% and 26%, respectively. Experimental validation with four-channel pTx showed 80% and 92% temperature reduction at the tips of wire 1 and wire 2, respectively. ConclusionThis pTx approach yields promising reductions in local SAR at the tips of unilateral and bilateral implanted wires while maintaining image quality in simulation and experiment. Magn Reson Med 78:2408-2415, 2017. (c) 2017 International Society for Magnetic Resonance in Medicine.
更多
查看译文
关键词
parallel RF transmission,deep brain stimulation,optimization,RF shimming
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要