Remote Electromagnetic Vibration of Steerable Needles for Imaging in Power Doppler Ultrasound.

2015 IEEE International Conference on Robotics and Automation (ICRA)(2015)

引用 13|浏览22
暂无评分
摘要
Robotic needle steering systems for minimally invasive medical procedures require complementary medical imaging systems to track the needles in real time. Ultrasound is a promising imaging modality because it offers relatively low-cost, real-time imaging of the needle. Previous methods applied vibration to the base of the needle using a voice coil actuator, in order to make the needle visible in power Doppler ultrasound. We propose a new method for needle tip vibration, using electromagnetic actuation of small permanent magnets placed inside the needle to improve needle tip visibility in power Doppler imaging. Robotic needle insertion experiments using artificial tissue and porcine liver showed that the electromagnetic tip vibration method can generate a stronger Doppler response compared to the previous base vibration method, resulting in better imaging at greater needle depth in tissue. It also eliminates previous issues with vibration damping along the shaft of the needle.
更多
查看译文
关键词
remote electromagnetic vibration,vibration damping,permanent magnet,electromagnetic actuation,medical imaging system,minimally invasive medical procedure,power Doppler ultrasound imaging,robotic needle steering system
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要