Response of Multi-strip Multi-gap Resistive Plate Chamber

JOURNAL OF INSTRUMENTATION(2015)

引用 3|浏览37
暂无评分
摘要
A prototype of Multi-strip Multi-gap Resistive Plate chamber (MMRPC) with active area 40 cm x 20 cm has been developed at SINP, Kolkata. Detailed response of the developed detector was studied with the pulsed electron beam from ELBE at Helmholtz-Zentrum Dresden-Rossendorf. In this report the response of SINP developed MMRPC with different controlling parameters is described in details. The obtained time resolution (sigma(t)) of the detector after slew correction was 91.5 +/- 3 ps. Position resolution measured along (sigma(x)) and across (sigma(y)) the strip was 2.8 +/- 0.6 cm and 0.58 cm, respectively. The measured absolute efficiency of the detector for minimum ionizing particle like electron was 95.8 +/- 1.3 %. Better timing resolution of the detector can be achieved by restricting the events to a single strip. The response of the detector was mainly in avalanche mode but a few percentage of streamer mode response was also observed. A comparison of the response of these two modes with trigger rate was studied.
更多
查看译文
关键词
Gaseous detectors,Timing detectors,Resistive-plate chambers,Instrumentation and methods for time-of-flight (TOF) spectroscopy