Medical and technical protocol for automatic navigation of a wireless device in the carotid artery of a living swine using a standard clinical MRI system.

MICCAI'07: Proceedings of the 10th international conference on Medical image computing and computer-assisted intervention - Volume Part I(2007)

引用 12|浏览36
暂无评分
摘要
A 1.5 mm magnetic sphere was navigated automatically inside the carotid artery of a living swine. The propulsion force, tracking and real-time capabilities of a Magnetic Resonance Imaging (MRI) system were integrated into a closed loop control platform. The sphere was released using an endovascular catheter approach. Specially developed software is responsible for the tracking, propulsion, event timing and closed loop position control in order to follow a 10 roundtrips preplanned trajectory on a distance of 5 cm inside the right carotid artery of the animal. Experimental protocol linking the technical aspects of this in vivo assay is presented. In the context of this demonstration, many challenges which provide insights about concrete issues of future nanomedical interventions and interventional platforms have been identified and addressed.
更多
查看译文
关键词
Magnetic Resonance Imaging,wireless,tracking,control,in vivo assay